GE Medical Dash 3000 Vital Signs Monitor View larger

GE Medical Dash 3000 Vital Signs Monitor

New product

GE Medical Dash 3000 Vital Signs Monitor Lightweight, durable, and ergonomic offer a high-performance, mobile monitoring solution that provides life-critical patient data no matter where you are

 

All GE Dash Monitors come complete with all accessories, manuals, quick start guide and 3 year factory warranty from GE Medical. Call or email us today for a quote on the GE Dash 3000 Series Monitors in a variety of configurations.

DASH 3000 is specifically designed to help you maintain monitoring continuity.

  • 8.4" Display, Transport with GE Ohmeda SpO2, stand alone, non-networked patient monitor.
  • English - U.S. 6ft. 1 (one) Lithium-Ion internally or externally rechargeable battery.
  • 4-Channel Integrated Thermal Printer.
  • Standard Handle.
  • Basic Software.
  • Typical accessories included, including Ops Manual & Service CD.
  • 2 year parts and labor warranty.

 

 

The GE Dash 3000 Monitor, Base model  features:

 

  • Color Display
  • 3-or 5-Lead ECG
  • Lethal Arythmia(V-Tach,V-Fib and Asystole)
  • Resp
  • Nellcor Sp02
  • Dinamap NIBP
  • Temp(1)
  • Recorder/Printer
  • Battery(1)
  • 1 Year Factory Warranty

 

 

SPECIFICATIONS

 

Display

Size:

Dash 3000 – 8.5 in.
Dash 4000 – 10.4 in. (diagonal)

Type:

Active-matrix color LCD

Resolution:

640 by 480 pixels

Number of Traces:

6 (maximum)

Number of seconds/trace

4.9 at 25 mm/sec

Sweep Speed:

6.25, 12.5, 25 mm/sec (with erase bar)

Controls

TrimKnob control
Five hard keys - Silence Alarm, Graph Go/Stop, NBP Go/Stop, Zero All and Power On/Off
Remote control option - Available

Alarms

Categories:

Patient status and system status

Priority:

4 levels – Crisis, Warning, Advisory and Message

Notification:

Audible and visual

Setting:

Default and individual

Silencing:

1 minute, current alarm only

Pause:

5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode and 5 minute, 15 minute, or permanent pause in OR mode

Volume:

Default 70 dB measured at 1 meter

Invasive Blood Pressure

Number of channels:

1 to 4 (optional); 3rd and 4th invasive lines only available with Masimo SpO2

Transducer sites

Arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atrial, intracranial and special

Transducer requirements:

Excitation voltage: 5 V dc ± 0.1%

Transducer output:

5 μV/V/mmHg

Input Specs

Range:

-25 mmHg to 300 mmHg

Offset:

± 150 mmHg

Output Specs

Frequency response:

dc to 50 Hz (-3 dB)

Zero balance range:

± 150 mmHg

Zero balance accuracy:

± 1 mmHg

Zero balance drift:

± 1 mmHg over 24 hours

Accuracy:

± 2% or ± 1 mmHg, whichever is greater (exclusive of transducer)

Alarms:

User-selectable upper and lower limits for systolic, diastolic and mean pressures

ECG

Standard leads available:

I, II, III, V, aVR, aVL and aVF

3, 5 and 10 leadwire cable:

I, II, III, V, aVR, aVL, aVF, V2, V3, V4, V5 and V6

Leads analyzed simultaneously:

I, II, III and V (multi-lead mode)

Lead fail:

Identifies failed lead

Alarms

User-selectable upper and lower heart rate limits

Input Specs

Voltage Range::

± 0.5 mV to ± 5 mV

Signal width:

40 ms to 120 ms (Q to S)

Heart rate range:

3 - 0 to 300 bpm

Input impedance:

Common mode - 10 MΩ at 50/60 Hz
Differential - 2.5 MΩ from dc to 60 Hz
Common mode rejection - 90 dB minimum at 50 or 60 Hz

Output Specs

Impulse response:

For an impulse of 3 mV applied for 100 ms
Displacement following impulse < 0.1 mV
Slope following impulse < 0.3 mV/s

Frequency response:

Response of non-permanent displays is limited by resolution to 40 Hz (-3 dB) @25 mm/s.
Specified upper frequency limits may vary by ± 2 Hz

Diagnostic mode:

0.67 Hz (+0.4 dB) to 100 Hz (-3 dB)

For compliance with China National Standard:

1.0 Hz (+0.4dB) to 75 Hz (-3 dB)

Monitoring mode:

0.67 (+0.4 dB) to 40 Hz (-3 dB)

Moderate mode:

0.67 (+0.4 dB) to 25 Hz (-3 dB)

Maximum mode:

5.0 Hz (-0.3 dB) to 25 Hz (-3 dB)

Noise:

30 μV (referred to input)

Pacemaker detection/rejection

Input voltage range:

± 2 mV to ± 700 mV

Input pulse width:

0.1 ms to 2 ms

Rise time:

10 μs to 100 μs

Over/under shoot:

2 mV (max)

Baseline drift:

< 0.5 mV per hour with a ± 700 mV, 2 ms

Pacemaker pulse:

Applied

Respiration

Measurement technique:

Impedance variation detection

Range:

0-200 breaths per minute for variations of 1.0 – 10.0 Ω

Respiration rate:

0-200 breaths per minute

Base impedance:

100-1000 Ω at 52.6 kHz

Detection sensitivity

0.4 to 10 Ω variation

Waveform display bandwidth

0.1 to 1.8 Hz (-3 dB)

Alarms:

User-selectable upper and lower respiration rate limits, and
user-selectable apnea limit

Temperature

Number of Channels:

2

Input Specs

Probe type:

YSI Series 400 or 700 (determined by input cable)

Temperature range:

0°C to 45°C (32°F to 113°F)

Resolution

± 0.1°C

Output Specs

Parameters displayed:

T1, T2

Accuracy (independent of source):

± 0.1°C for YSI series 400

Probes:

± 0.3°C for YSI series 700 probes

Alarms:

User-selectable upper and lower limits for T1, T2

Cardiac Output

Input Specs

Probe type:

In-line or bath probe

Catheter size:

5F, 6F, 7F, 7.5F and 8F

Injectate volume

3, 5 or 10 cc

Output Specs

Parameters displayed:

Cardiac output, blood temperature, injectate temperature and trial number

Range

Cardiac output:

0.2-15 (liters per minute)

Blood temperature:

30-42°C

Injectate temperature:

0-30°C

Accuracy

Cardiac output:

± 5%

Blood temperature:

± 0.2°C

Injectate temperature:

± 0.3°C

Frequency response:

dc to 15 Hz ± 2 Hz

Pulse Oximetry

Parameters monitored:

Arterial oxygen saturation (SpO2) and peripheral pulse rate (PPR)

SpO2 range:

Nellcor-compatible 50-100%
Masimo 30-100%

PPR range:

Nellcor-compatible 30-300 BPM; Masimo 25-240 BPM

Accuracy:

Actual accuracy depends on probe. Please reference manufacturer's specifications.

Nellcor compatible SpO2:

± 2% (70-100% SpO2); ± 3% (50-69% SpO2)

Masimo SpO2:

± 2% Adults/Pediatric (70-100% SpO2); SpO2 ± 3% Neonates ≤ 69% unspecified

PPR:

± 3 beats per minute

Alarms:

User-selectable upper and lower limits for SpO2 and PPR

DINAMAP Non-Invasive Blood Pressure

Technology:

DINAMAP classic and SuperSTAT (SuperSTAT only available with Masimo SpO2)

Measurement technique:

Oscillometric

Displayed parameters:

Systolic, diastolic and mean pressures, time of last measurement

Measurement modes:

Adult ICU and OR modes; manual, auto and stat, neonatal mode; manual and auto

  • Systolic
    Adult - 30-285 mmHg
    Pediatric - 30-235 mmHg
    Neonate - 30-140 mmHg
  • Map
    Adult - 20-260 mmHg
    Pediatric - 20-220 mmHg
    Neonate - 20-125 mmHg
  • Diastolic
    Adult - 10-220 mmHg
    Pediatric - 10-210 mmHg
    Neonate - 10-110 mmHg

Pulse rate, as displayed in tabular trends:

  • Adult - 30-200 bpm
  • Pediatric - 30-200 bpm
  • Neonate - 30-220 bpm

Other Specs

Overall system accuracy:

Meets or exceeds SP 10-1992 AAMI standards

Automatic cycle times:

0-8 hours

Tubing length:

12 feet adult
8 feet neonatal

Automatic cuff deflation:

Cycle time exceeding 3 minutes (90 seconds neonatal), French mode – Cycle time exceeding 2 minutes (60 secondsneonatal), power off, or cuff pressure exceeds 294 mmHg (± 6 mmHg) adult, 250 (± 5 mmHg) pediatric, 147 (± 3 mmHg) neonatal

Cuff sizes:

Thigh, large adult, adult, small adult, child, infant and neonatal, sizes #5 - #1 and assorted long sizes

Alarms:

User-selectable upper and lower limits for systolic, diastolic and mean pressures

CO2

Technology

Supports Novametrix CapnoStat (mainstream) and LoFlo (low-flow sidestream) CO2 technologies

Principle of operation:

Non-dispersive infrared (NDIR) single beam optics, dual wavelength and no moving parts

Warm-up time:

2 minutes warm-up time to meet accuracy specifications; waveform immediate upon power up, calculated end tidal after two breaths

Cable length (mainstream):

8 foot (2.4 m)

Sample line length (low-flow sidestream):

7 foot (2.1 m)

Information Displayed:

Inspired and expired CO2 concentrations in %, mmHg or kPa; respiratory rate, continuous CO2 waveform

Measurement range (at 760 mmHg at an ambient temperature of 25°C):

0-100 mmHg, 0-13%, 0-12.5 kPa
PiCO2 /FiCO2 - 0-50 mmHg, 0-6.5%, 0-6.25 kPa

Respiration rate range:

Low-flow SS 0-150 breaths/min

Mainstream:

0-120 breaths/min

Accuracy (at 760 mmHg at an ambient temperature of 25°C)
MS:

±2 mmHg or 5%, whichever is greater
SS 0-40 mmHg ± 2 mmHg; 41-70 mmHg ± 5% of reading; 71-100 mmHg ± 8% of reading; all specifications ± 12% of actual from 80-150 BrPM

Display resolution:

1 mmHg

Rise time:

Less than 200 ms (low-flow sidestream); less than 60 ms (mainstream adult reusable or SPU); less than 50 ms (mainstream infant reusable or SPU)

Respiration rate accuracy:

± 1 breath/min

Compensations

Automatic barometric pressure:

± 25 mmHg from 530-785 mmHg

 

Operator-selectable O2/N2O compensation

Calibration

Mainstream:

No routine user calibration required
15 second airway adapter zero performed when changing to a different style of airway adapter

Low-flow sidestream

No routine user calibration required

Airway adapters and sample lines – mainstream (airway adapters)

Types:

Adult reusable (standard), adult disposable, infant

Deadspace:

Adult reusable/disposable < 5 cc
Infant disposable < 1 cc
Taper meets ISO 5356-1

Low-flow sidestream airway adapters

Types:

Adult reusable (standard), adult disposable, infant

Deadspace:

Adult reusable/disposable < 7.3 cc
Infant disposable < 1 cc

Low-flow sidestream sample lines

Adult, pediatric and infant:

Nasal CO2 and nasal CO2/O2

Adult and pediatric:

Nasal/oral CO2 and nasal/oral CO2/O2

Alarms

CO2:

High inspired CO2; high/low expired CO2

Respiratory rate:

Adjustable high and low

Paper Recorder

Method:

Thermal dot array

Horizontal resolution:

480 dots/in. @25 mm/sec.

Vertical resolution:

200 dots/in.

Number of waveform channels:

four

Paper width:

50 mm (1.97 in.)

Paper length:

30 m (100 ft.)

Paper speed:

0.1, 0.5, 1, 5, 10, 12.5, 25 and 50 mm/sec. (± 2%)

Analog Output

ECG

Gain:

1 V/mV ± 10%

DC offset:

± 100 mV (max)

Noise:

< 5 mV peak to peak 0-300 Hz

Frequency response:

Refer to Frequency Response section under ECG

Blood pressure

Gain:

10 mV/mmHg ± 2%

DC offset:

± 20 mV (max)

Noise:

< 5 mV peak to peak 0-300 Hz

Frequency response:

dc to 50 Hz-0/+2 Hz

Battery

Battery type:

Exchangeable Lithium-Ion

Maximum number of batteries:

2

Voltage:

11.1 V (nominal)

Capacity:

≥ 3.45 Ah (varies with manufacturers)

Charge time:

Less than 4 hours each

Run time:

4 to 5 hours

Battery life:

500 cycles to 50% capacity

Wireless LAN

Operating frequency:

2.4 to 2.5 GHz

Transmit power:

100 mW

Data rate:

1Mbps and 2Mbps per channel

Radio technology:

Frequency-hopping spread spectrum

Communication protocol:

IEEE 802.11

 

UL 1950 Listed (ITE 9B97), CE Mark RF Standard
US - FCC Part 15 Class B, Europe: ETS 300 328 and ETS 300 826

Environmental Specifications

Power requirements

90-132 VAC 50/60 Hz 2.0A
190-264 VAC 50/60 Hz 1.0A

Power consumption:

75 W (fully loaded)

Cooling:

Convection

Heat dissipation:

240 Btu/hr. (max)

Operating Conditions

Ambient temperature:

0-40°C (32-104°F)

While charging batteries:

0-35°C (32-95°F)

CO2 sensor:

10-40°C (50-104°F)

Relative humidity:

5-95% @40°C

Vibration:

MIL-STD 810E, Method 514.4, Category 1

Altitude:

-273 m to 2,943 m (-896 to 9,655 ft.)

Storage Conditions

Do not exceed
Maximum - 70°C (158°F) at 95% relative humidity
Minimum - -40°C (-40°F)

CO2 sensor:

-30 to 65°C (-22 to 149°F)

Batteries:

-20 to 60°C (-4 to 140°F)

Physical Specifications

Height:

26 cm (10.25 in.)

Depth:

20 cm (8 in.)

Width:

28 cm (11.0 in.)

Weight:

5.08 kg (11.2 lbs.)

Warranty

Standard warranty is 3 years.